Parameter |
Default Setting |
Description |
Date |
dd mmm yyyy |
|
Time |
hh:mm |
24 hour clock |
Print on Mark |
On |
Prints a 3 second strip when “Mark” is pressed. |
Print on Charge |
On |
Prints a continuous strip during charging until a shock is
delivered, device is disarmed, or “Strip” is pressed. |
Print on Shock |
On |
Prints a 12 second strip when a shock is delivered. |
Print on Alarm |
On |
Prints a 6 second strip during alarms. |
Printer Delay |
6 Sec Delay |
Prints 6 seconds of information present prior to printer activation. |
Pace Pulse Markers |
Leads & Pads |
Shows pace pulse markers on the ECG when an internal pacemaker
is detected. |
ECG Lead Cable |
3-Lead |
Selects monitoring electrodes source. |
QRS Beeper |
On |
Audible beep with each QRS. |
Power on Lead |
Paddle Mode |
|
Parameter |
Default Setting |
Description |
AED Shock Series |
1 |
Maximum number of shocks before prompting to check patient,
pulse, and begin CPR. |
Shock Series Timer |
60 Sec |
Defines the number of seconds that must pass before the next
shock is considered the first shock of a new shock series, rather
than the next shock of the current shock series. |
Device Initiated Analysis |
On |
Initiates ECG analysis when the Heartstream XL is turned on
in AED mode for new use. |
Automatic Re-Analysis |
On |
Initiates ECG analysis in between shocks within a shock series. |
Rhythm Monitoring |
On |
Monitors the ECG for potentially shockable rhythms when the
Heartstream XL is not analyzing, defibrillating, or paused. |
“Check Patient” Timer |
30 Sec |
Defines how often (in seconds) the “Check Patient” prompt
is repeated when Rhythm Monitoring detects a potentially shockable
rhythm. |
European Protocol |
Off |
For European market needs. |
“Pause” Timer |
120 Sec |
Defines the duration of the pause time (in seconds) when “Pause”
is pressed. |
Lead Select |
On |
Turns on ECG monitoring via ECG leads. |